Xoran MiniCAT System for ENT (server and work station), Xoran Technologies, Ann Arbor, MI. Cone beam CT for sinus, temporal bones, and maxillofacial imaging.

This recall has been terminated (originally issued December 23, 2009).

Company
Xoran Technologies, Inc.
Recall Initiated
September 17, 2008
Posted
December 23, 2009
Terminated
December 23, 2009
Recall Number
Z-0561-2010
Quantity
82
Firm Location
Ann Arbor, MI
Official Source
View on FDA website ↗

Reason for Recall

The system may overheat, causing poor image quality, interruption in image acquisition and/or damage to the system. Also, if the system is not allowed to cycle down prior to being restarted after an emergency stop, the x-ray and gantry motions will both resume.

Distribution

Nationwide Distribution

Lot / Code Info

All units.

Root Cause

Device Design

Action Taken

A letter dated 9/17/08 was sent to each account telling them that the PLC would be upgraded. The firm visited each account and made corrections.