RecallDepth

One Touch SureSoft Lancing Device in carton. Part Number 021-140-01, aka OT SureSoft Hospital, Gentle 200

Company
Lifescan Inc
Recall Initiated
March 30, 2007
Posted
August 16, 2007
Terminated
December 3, 2007
Recall Number
Z-1165-2007
Quantity
632 cartons (200 units per carton)
Firm Location
Milpitas, CA

Reason for Recall

Lancet tip may not fully retract and cause needle stick and exposure to used lancets.

Distribution

Product was distributed nationwide to distributors and wholesale dealers.

Lot / Code Info

Lot SA4426; Part number 021-140-01

Root Cause

Nonconforming Material/Component

Action Taken

The firm has conducted a health hazard evaluation, and has initiated a corrective action plan. The firm initated recall on March 30, 2007, and notifications were sent to hospital end users known to Lifescan and its direct accounts. In addition, a customer service script has been developed to address any questions from incoming users. The firm has requested direct accounts to perform subrecalls on Lifescan's behalf to notify customers of the recall as needed.

More recalls by Lifescan Inc

View all recalls by this company →