OneTouch Delica Lancing Device; LifeScan Inc., Milpitas, CA Testing of Blood Glucose Levels
- Company
- Lifescan Inc
- Recall Initiated
- May 27, 2011
- Posted
- July 15, 2011
- Terminated
- January 5, 2012
- Recall Number
- Z-2835-2011
- Quantity
- 1,032,515 in US
- Firm Location
- Milpitas, CA
Reason for Recall
Individuals attempting to remove a lancet from the OneTouch Delica lancing device without following instructions for use may experience an accidental needle stick.
Distribution
Worldwide Distribution - USA (nationwide) and Canada.
Lot / Code Info
Model Number 022-137-01
Root Cause
Other
Action Taken
Lifescan sent an Urgent Medical Device Correction letter dated May 27, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were told that no action on their part is needed and were asked not to return any of the affected product. Customers were instructed to continue to distribute the product. Healthcare professionals and pharmacies were asked to share the information with their patients who use the OneTouch Delica Lancing Device. Patients with questions should contact Customer Care Line at 1-800-574-6138. For questions regarding this recall call 1-800-600-7226.