LifeScan brand OneTouch Ultra Control Solution, Part Numbers/Model Numbers: 01045808, 02141603. Testing of Blood Glucose Levels.

Company
Lifescan Inc
Recall Initiated
February 25, 2013
Posted
April 17, 2013
Terminated
October 23, 2013
Recall Number
Z-1129-2013
Quantity
132,301 units total
Firm Location
Milpitas, CA

Reason for Recall

When OneTouch Ultra and OneTouch Select test strips are tested with Ultra and Select control solution the meter may display a result outside the control solution test range printed on the strip vial label even though the system is working properly.

Distribution

Worldwide Distribution-USA (nationwide) including DC and Puerto Rico and the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, 10, IL, IN, KS, KY, LA, MA, MD, ME, Ml, MN, MO, MS, MT, NC, NO, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SO, TN, TX, UT, VA, VI, VT, WA, WI,WV, and WY, and the countries of Russia and Ukraine.

Lot / Code Info

2AA2G01, Exp. Date: 7-31-2014; 2AA2G02, Exp. Date: 7-31-2014; 2AA2G03, Exp. Date: 8-31-2014; 3AA2G04, Exp. Date: 12-31-2014; 3AA2G05, Exp. Date: 12-31-2014

Root Cause

Nonconforming Material/Component

Action Taken

LifeScan sent an "URGENT MEDICAL DEVICE RECALL" letter dated February 25, 2013 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. Customers may contact OneTouch Healthcare Professional Line at 866-461-0089 for questions about this notice.