Hitachi Echelon Magnetic Resonance Imaging System, Serial Number: V-001, equipped with CTL coil type: MR-CTL-150.

This recall has been terminated (originally issued November 4, 2006).

Company
Hitachi Medical Systems America Inc
Recall Initiated
August 14, 2006
Posted
November 4, 2006
Terminated
September 19, 2012
Recall Number
Z-0152-2007
Quantity
1 unit
Firm Location
Twinsburg, OH
Official Source
View on FDA website ↗

Reason for Recall

Component defect/overheating device; A malfunctioning decoupling circuit in the CTL coil caused high current flow and circuit overheating.

Distribution

The recalled unit was distributed to a single customer.

Lot / Code Info

Echelon unit Serial Number: V001. CTL coil component type : MR-CTL-150, N/N KH19636602.

Root Cause

Other

Action Taken

The recalling firm issued a Device Correction letter to the clinic customer on 9/15/2006 making them aware of this potential problem and of their completed corrective action on 8/14/2006, by a Service Rep visit.