UST-2265-2 Probe, Part Number: UST-2265-2 - Product Usage: The UST-2265-2 probe is intended to be used for diagnostic ultrasound imaging or fluid flow analysis of the human body as Continuous-wave Doppler (CWD) for Cardiac Adult, Cardiac Pediatric, and Peripheral Vessel.

Company
Hitachi Medical Systems America Inc
Recall Initiated
January 31, 2020
Terminated
July 31, 2020
Recall Number
Z-1341-2020
Quantity
18
Firm Location
Twinsburg, OH

Reason for Recall

These 18 probes may have mis-wiring in the inner cables, causing decreased sensitivity and thus resulting in lower peak velocity readings.

Distribution

Domestic distribution to CA, FL, GA, MA, NV, OH, TX, and WV. Foreign distribution to Canada.

Lot / Code Info

Serial Number: G3007689, G3007688, G3007687, 206H2628, 206H2629, 206H2630, 206H2620, 206H2621, 206H2622, 206H2623, 206H26224, 206H2625, 206H2626, 206H2627, 205X3897, 205X3898, 205X3899, and 205X3900

Root Cause

Process change control

Action Taken

On January 31, 2020, the firm initiated customer communications via Urgent Medical Device Recall letters. Customers were informed of the potential mis-wiring issue. Clinicians are advised to discontinue use of the probes to minimize adverse events. Hitachi Service will contact each affected site to schedule a service visit. Hitachi will test all affected probes and remove and replace any probes that do not pass the test. The correction is expected to begin February 2020.