Hitachi Scenaria Whole-body X-ray CT System

Company
Hitachi Medical Systems America Inc
Recall Initiated
October 16, 2019
Terminated
April 30, 2020
Recall Number
Z-0271-2020
Quantity
39
Firm Location
Twinsburg, OH

Reason for Recall

There is a possibility that the cables to the controller unit may not be affixed properly, creating the potential for a cable to detach during scanning and eject from the system.

Distribution

The devices were distributed to the following US states: CA, FL, GA, IA, IN, MD, MT, NC, NY, OH, OK, SD, TN, TX, WY, and PR. The products were distributed to the following foreign countries: Brazil and Mexico.

Lot / Code Info

Scenaria units S5002-S5014, S5017-S5044

Root Cause

Nonconforming Material/Component

Action Taken

An Urgent Medical Device Correction Notice dated 10/15/19 was sent to customers. Hitachi will inspect all heat controller units for Scenaria systems S5002-S5014 and S5017-S5044 to ensure that the cables and cable loop clamp are affixed as instructed and apply modifications as needed. We expect to begin this process in late October I early November 2019. This action will be implemented by Hitachi Service at no charge to you. Your Service Area Manager will contact you for scheduling. Please contact me if you need further information regarding this report. You can send any questions, comments or requests to recalls@hitachihealthcare.com.