AEROSET/ARCHITECT c8000 Urea Nitrogen Models: 7D75-01; 7D75-20; 7D75-30
This recall has been terminated (originally issued December 18, 2003).
- Recall Initiated
- November 24, 2003
- Posted
- December 18, 2003
- Terminated
- January 25, 2012
- Recall Number
- Z-0240-04
- Quantity
- All lots.
- Firm Location
- South Pasadena, CA
- Official Source
- View on FDA website ↗
Reason for Recall
False low results may occur with out error codes due to substrate depletion in samples with more than 220mg/dL BUN.
Distribution
Nationwide and Mexico, Brazil, Germany, Singapore, Grand Cayman, Venezuela, Puerto Rico, Hong Kong, Australia, Columbia, Dominican Republic, China, Trinidad, Argentina, Canada, New Zealand, Bahamas
Lot / Code Info
All lots.
Root Cause
Other
Action Taken
Firm sent letter to customers dated November 24,2003 which includes a new package insert to use with this reagent. Instructions were changed to correct low resulting situation. Subrecall was requested.