AEROSET/ARCHITECT c8000 Urea Nitrogen Models: 7D75-01; 7D75-20; 7D75-30

This recall has been terminated (originally issued December 18, 2003).

Company
Abbott Laboratories Diagnostic Div
Recall Initiated
November 24, 2003
Posted
December 18, 2003
Terminated
January 25, 2012
Recall Number
Z-0240-04
Quantity
All lots.
Firm Location
South Pasadena, CA
Official Source
View on FDA website ↗

Reason for Recall

False low results may occur with out error codes due to substrate depletion in samples with more than 220mg/dL BUN.

Distribution

Nationwide and Mexico, Brazil, Germany, Singapore, Grand Cayman, Venezuela, Puerto Rico, Hong Kong, Australia, Columbia, Dominican Republic, China, Trinidad, Argentina, Canada, New Zealand, Bahamas

Lot / Code Info

All lots.

Root Cause

Other

Action Taken

Firm sent letter to customers dated November 24,2003 which includes a new package insert to use with this reagent. Instructions were changed to correct low resulting situation. Subrecall was requested.