Abbott Clinical Chemistry Bilirubin Calibrator, List # 1E66-03; Lot # 20881M100

This recall has been terminated (originally issued August 24, 2005).

Company
Abbott Laboratories Diagnostic Div
Recall Initiated
August 5, 2005
Posted
August 24, 2005
Terminated
March 7, 2012
Recall Number
Z-1400-05
Quantity
213
Firm Location
South Pasadena, CA
Official Source
View on FDA website ↗

Reason for Recall

The product issue is that the value sheet for the Clinical Chemistry Bilirubin Calibrator, lot #20881M100 calibrator value sheet, has the calibrator values transposed for the Total Bilirubin Assay LN 7D60, Total Bilirubin Assay LN 8G62, and Neonatal Bilirubin Assay LN 9D88.

Distribution

Nationwide and Argentian, Brazil, Canada, Cayman, Chile, Columbia, Germany, Hong Kong, Jamaica, New Zealand, Puerto Rico, Singapore, Thailand, Venzuela

Lot / Code Info

List # 1E66-03; Lot #20881M100

Root Cause

Other

Action Taken

A Device Correction letter correcting the transposition of the calibrator values dated August 5, 2005, was mailed to all customers receiving the Clinical Chemistry Bilirubin Calibrator, LN 1E66-03, Lot Number 20881M100.