Clinical Chemistry Apolipoprotein A1, List Number 9D92-20

This recall has been terminated (originally issued January 11, 2006).

Company
Abbott Laboratories Diagnostic Div
Recall Initiated
November 10, 2005
Posted
January 11, 2006
Terminated
March 20, 2007
Recall Number
Z-0383-06
Quantity
617 (134 in US)
Firm Location
South Pasadena, CA
Official Source
View on FDA website ↗

Reason for Recall

These specific protein assays are not meeting the 57-day on board stability claim

Distribution

Nationwide, plus Australia, Canada, Columbia, Germany, Guatemala, Hong Kong, mexico, & New Zealand

Lot / Code Info

Lot Number(s): 30029HW00, 28002HW00, 26027HW00, & 23006HW00

Root Cause

Other

Action Taken

A Device Correction letter identifying the changes being made to the frequency of QC monitoring for these assays, will be distributed to all AEROSET and ARCHITECT c8000 customers which utilize the IgM, Prealbumin and Apolipoprotein Al Specifc Protein assays on November 10,2005 via direct mailing. Outside the U.S. each country organization will need to determine the appropriate method for the effectiveness check..