Hitachi ECLOS Computed Tomography (CT) Scanners, Software Version: 2.11 or less.

This recall has been terminated (originally issued July 7, 2009).

Company
Hitachi Medical Systems America Inc
Recall Initiated
April 14, 2009
Posted
July 7, 2009
Terminated
September 21, 2012
Recall Number
Z-1652-2009
Quantity
4 units
Firm Location
Twinsburg, OH
Official Source
View on FDA website ↗

Reason for Recall

Incorrect scale on image: A software error can occur if two (2) different Field-of-View (FOV) settings are registered in a protocol using Multi-Reconstruction. This could cause the image generated to be displayed with the wrong scale factor rendering anatomical measurements incorrect.

Distribution

Nationwide distribution, including Puerto Rico (PR).

Lot / Code Info

The recalled ECLOS system units are identified with the following Code Numbers: E1014, E5003 through E5005

Root Cause

Software design

Action Taken

HMSA issued an URGENT - DEVICE CORRECTION notification to their customers on 4/14/2009. The notification describes the potential software error and the condition under which these image distortions can occur. The notification goes on to provide the customer with a work-around procedure which they can use to avoid encountering this issue with their current software configuration(s). The letter states that the firm will be installing updated software on all affected systems beginning in June, 2009, and that a Hitachi Service Rep. will be contacting each customer and making arrangements to come out and install a software upgrade on the affected system(s) at no cost to the customer. Lastly, the letter provides a telephone # to the firm's Applications Helpline: 1-877-228-2777, which the customer can use to obtain additional information, if desired.