NeoPAP Neonatal CPAP/Humidification System (Ventilator, Continuous). Model number 1025310
- Company
- Respironics, Inc.
- Recall Initiated
- September 19, 2005
- Posted
- July 1, 2006
- Terminated
- July 11, 2006
- Recall Number
- Z-1191-06
- Quantity
- 232 units
- Firm Location
- Murrysville, PA
Reason for Recall
Gas input pressures over 62 psi may cause CPAP pressure oscillation
Distribution
World wide (including medical facilities in AZ, CA, CO, GA, IA, IL, MA, MD, MN, MO, NC, NE, NH, NJ, OH, OK, OR, PA, PR, TN, TX, VA, WI, and WV. Japan, Ireland, Great Britain, Sweden, Portugal, France, Chile, Argentina, Hong Kong, South Africa, Greece, Egypt, Saudia Arabia, Norway, Spain, Belgium and Canada
Lot / Code Info
Serial numbers 5310050616-01, 5310050616-02, 5310050621-01
Root Cause
Other
Action Taken
The recalling firm issued a recall letter to their customers dated 9/19/05 informing them of the problem and the corrective action. (fitting unit with external 50 psi pressure regulator for both air and oxygen)