Trilogy 100 Ventilator, Model/Catalog Number 1054260, U1054260 (Domestic). Philips Respironics, Monroeville, PA 15146. Provides continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation.

Company
Respironics, Inc.
Recall Initiated
June 10, 2010
Posted
June 24, 2010
Terminated
November 18, 2010
Recall Number
Z-1896-2010
Quantity
23 units
Firm Location
Murrysville, PA

Reason for Recall

It is possible that the battery could electrically short, resulting in a potential for fire or a thermal event.

Distribution

Worldwide Distribution -- United States (MT, IL, IA, NY, NC, TX, CA and FL), France, South Korea and Turkey.

Lot / Code Info

Serial Numbers: TV110051146, TV110051147, TV110051347, TV110051335, TV110050717, TV110052245, TV110051033, TV110051130, TV110050733, TV110051133, TV110051336, TV110051144, TV110051241, TV110051236, TV110051344, TV110051333, TV110051338, TV110051340, TV110051352, TV110051348

Root Cause

Nonconforming Material/Component

Action Taken

Philips Respironics has contacted all of its consignees by phone and instructed them to remove the potentially affected external battery packs from their inventory and any that may be on their patients/end users. These consignees have been instructed to quarantine batteries they have removed away from flammable materials. Product return/replacement instructions were provided. For further information, contact Philips Respironics at 1-877-387-3311.