Power Board Kit for use with the PLV-100 Portable Lifecare repairs

This recall has been terminated (originally issued September 18, 2008).

Company
Respironics, Inc.
Recall Initiated
July 8, 2008
Posted
September 18, 2008
Terminated
April 23, 2009
Recall Number
Z-2345-2008
Quantity
197
Firm Location
Murrysville, PA
Official Source
View on FDA website ↗

Reason for Recall

Certain PLV-100 ventilators may not trigger a signal to activate third-party remote alarms or nurse call systems upon device failure.

Distribution

Worldwide Distribution ---USA: NC, IL, NJ, MO, TX , OH, IA, and OK and countries of Japan, Denmark, Korea, Australia, Finland, Canada, Argentina, New Zealand, Hong Kong, and India.

Lot / Code Info

Model/Catalog number PLV-100. Circuit board part numbers 1034786 version 5 or 1036038 version 3

Root Cause

Finished device change control

Action Taken

The recalling firm issued an URGENT MEDICAL DEVICE RECALL letter dated 7/8/08. The letter informed the customer of the problem and the need to contact the firm for a replacement. Contact Respironics Customer Service at 1-877-387-331 for assistance.