RecallDepth

Product is SoftPath ASCII Software Releases 1.2, 2.1, 2.2, and 2.3.

This recall has been terminated (originally issued September 22, 2008).

Company
Scc Soft Computer
Recall Initiated
February 23, 2006
Posted
September 22, 2008
Terminated
November 20, 2008
Recall Number
Z-2463-2008
Quantity
201
Firm Location
Clearwater, FL
Official Source
View on FDA website ↗

Reason for Recall

In the creation of revised report and supplemental reports diagnosis, text was inserted from another case.

Distribution

Distributed to 106 consignees located throughout the United States and Canada.

Lot / Code Info

Version 1.2, 2.1, 2.2 and 2.3

Root Cause

Software design

Action Taken

Consignees were notified of recall and provided with a software correction patch ICC number 3618.