Product is SoftPath ASCII Software Releases 1.2, 2.1, 2.2, and 2.3.
This recall has been terminated (originally issued September 22, 2008).
- Company
- Scc Soft Computer
- Recall Initiated
- February 23, 2006
- Posted
- September 22, 2008
- Terminated
- November 20, 2008
- Recall Number
- Z-2463-2008
- Quantity
- 201
- Firm Location
- Clearwater, FL
- Official Source
- View on FDA website ↗
Reason for Recall
In the creation of revised report and supplemental reports diagnosis, text was inserted from another case.
Distribution
Distributed to 106 consignees located throughout the United States and Canada.
Lot / Code Info
Version 1.2, 2.1, 2.2 and 2.3
Root Cause
Software design
Action Taken
Consignees were notified of recall and provided with a software correction patch ICC number 3618.