Impac Medical Systems Sequencer Product Usage: The intended us of Sequencer is to assists in the process of patient treatment on a radiotherapy treatment machine. Sequencer provides the capability to notify the user of actions that need to take place prior to treatment, to display reference images for set up purposed and to auto setup the machine to predefined settings. The actual settings are read from the treatment machine, via the machines communication interface, and compared to predefined values. if a mismatch occurs between the planned values and the actual machine settings, treatment is prohibited. At appropriate points during the treatment, the actual delivered values are recorded to provide treatment tracking.

This recall has been terminated (originally issued October 17, 2011).

Company
Impac Medical Systems Inc
Recall Initiated
September 9, 2011
Posted
October 17, 2011
Terminated
March 6, 2012
Recall Number
Z-0052-2012
Quantity
11 devices
Firm Location
Sunnyvale, CA
Official Source
View on FDA website ↗

Reason for Recall

Varian reported a problem related to certain collimator rotation values which are incorrect on TrueBeam compared to planned values in Mosaiq.

Distribution

Nationwide Distribution (USA) - including the states of: AZ, CA, FL, NJ, NY, NV, TX, VA and WA.

Lot / Code Info

All sites with the configuration of the 4D IEC 1217 source file used for the Varian TrueBEam treatment machine.

Root Cause

Nonconforming Material/Component

Action Taken

" Impac Medical Systems Inc., sent an USER NOTICE letter to all affected consignees on September 20, 2011. The letters identified the source characterization error, clinical impact, and final resolution. The letter states that Elekta Support will follow up with each consignee and remotely correct the source characterization. Customers were instructed to sign, date and return the attached Confirmation of Receipt Form by one of the three ways indicated. The letter instructs consignees to distribute this notice to any and all users of IMPAC software that may potentially be affected by the issue. For questions, concerns, and requests for upgrade, email support@impac.com or call. The letter provided consignees with a complete list of contact information..