SEQUENCER for Mosaiq interfaced to Electa SL75-5 machines. Sequencer Verify and Record System is to assist in the process of patient treatment on a radiotherapy treatment machine with software control.

This recall has been terminated (originally issued February 4, 2011).

Company
Impac Medical Systems Inc
Recall Initiated
March 10, 2010
Posted
February 4, 2011
Terminated
April 6, 2011
Recall Number
Z-1029-2011
Quantity
40
Firm Location
Sunnyvale, CA
Official Source
View on FDA website ↗

Reason for Recall

Computer interface is defected. Computer based verification system for recording patient treatment on a radiotherapy treatment machine fails to record treated field.

Distribution

Worldwide Distribution -- US, Canada, Germany, and France.

Lot / Code Info

Sequencer for MOSAIQ versions: 1.20, 1.30 ; 1.50 , 1.60 and 2.00, interfacing to an Elekta SL 75-5 Linacs.

Root Cause

Software design

Action Taken

A User Notice was sent by the firm to its consignees. It identified the affected product and the reason for the recall. It also discussed the clinical impact and interim mitigations. The firm states in the letter that a fix for the problem and been developed and is available. Customers are to contact IMPAC Support for the patch to the SL75-5 interface. If customers have questions, concerns, or requests, they can email support@impac.com or one of the numbers listed in the letter for the applicable country. The letter should be distributed to all users affected by the issue at the customer's organization.