Elekta Impac Software Sequencer Verify & Record System, for MOSAIQ, manufactured by Elekta Impac Software, Sunnyvale, CA. The intended use of Sequencer (generically referred to as a verification system) is to assist in the process of process of patient treatment on a radiotherapy treatment machine.

This recall has been terminated (originally issued October 28, 2010).

Company
Impac Medical Systems Inc
Recall Initiated
February 19, 2010
Posted
October 28, 2010
Terminated
March 28, 2011
Recall Number
Z-0159-2011
Quantity
19 units.
Firm Location
Sunnyvale, CA
Official Source
View on FDA website ↗

Reason for Recall

Mistreatment-- The database conversion utility used during the upgrade errantly changes the start gantry angle to zero. Any affected treatment fields, if not already treated, would start delivering from zero instead of the original prescribed value.

Distribution

Worldwide Distribution -- USA, Australia, Belgium, France, Germany, Hong Kong, Ireland, Israel, and Italy

Lot / Code Info

Sequencer for MOSAIQ 1.60W3 (including all service packs), 2.00T2 (inclusing all service packs), 2.10L5 (including all service packs) and 2.10M7 (including all service packs).

Root Cause

Process change control

Action Taken

ELEKTA IMPAC SOFTWARE sent a USER NOTICE dated February 19, 2010, to all customers. The notice identified the product, the problem, and the action to be taken by the customers. Customers were urged to verify the start angle of the gantry against the treatment field definition prior to delivery. Review all treatment plans and charts with conformal arcs. If a treatment plan with an errant gantry start angle was identified, customers were to cease using that plan immediately. A workaround was to be re-imported into their current version of MOSAIQ. IMPAC Support was to follow-up with each site and execute a script to detect and correct all affected treatment plans. For questions, concerns, and requests for upgrade, customers were to email support@impac.com or call their support center.