Monaco Radiation Treatment Planning Workstation. Monaco Release 2.03.00 and above. The Monaco RTP system is used to create treatment plans for any cancer patient for whom external beam intensity modulated radiation therapy (IMRT) has been prescribed.

This recall has been terminated (originally issued June 14, 2011).

Company
Computerized Medical Systems Inc
Recall Initiated
January 13, 2011
Posted
June 14, 2011
Terminated
April 3, 2014
Recall Number
Z-2561-2011
Quantity
94
Firm Location
Maryland Heights, MO
Official Source
View on FDA website ↗

Reason for Recall

Monaco: For VMAT plans, the measured dose of radiation delivered is lower than the software calculated dose by about 5% in Monaco Release 2.03.00 and 2.04.00.

Distribution

Worldwide Distribution-- USA (nationwide) and countries of Austria, Australia, Belarus, Canada, China, Cyprus, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, India, Israel, Italy, Japan, Mexico, Netherlands, Poland, Singapore, Slovenia, South Korea, Spain, Switzerland, Taiwan, Thailand and UK.

Lot / Code Info

Release 2.03.00 and above.

Root Cause

Software design

Action Taken

Computerized Medical Systems, Inc. sent a "USER NOTICE" on January 13, 2011, to all affected users. This notice identified the product, the problem, and the action needed to be taken by the customer. A return postcard is included for the customer to confirm receipt of the user notification. The firm intends for correction in the dose calulation engine to appear in the Monaco Release 3.00 in March 2011. Any further questions please call (403) 830-8023.

More recalls by Computerized Medical Systems Inc