ERGO++ Stereotactic Radiation Treatment Planning System, ERGO++ Release 1.6.3 and 1.6.3.1 Product Usage: Used to create treatment plans for any cancer patient for who external beam radiation therapy has been prescribed. It is an accessory to linear accelerators used for radiation therapy. It is indicated for use in the planning of 3-dimensional radiation therapy.

This recall has been terminated (originally issued June 23, 2011).

Company
Computerized Medical Systems Inc
Recall Initiated
December 11, 2009
Posted
June 23, 2011
Terminated
April 1, 2014
Recall Number
Z-2645-2011
Quantity
45
Firm Location
Maryland Heights, MO
Official Source
View on FDA website ↗

Reason for Recall

ERGO Release 1.6.3 is overestimating the MU values.

Distribution

Nationwide distribution to medical facilities in CA, MO, NC, ND, NV, NY, OH, OK, TX, and WA. Foreign distribution was made to Brazil, Cuba, Czech Republic, France, Germany, India, Italy, Japan, Malaysia, Poland, Portugal, Serbia, and Spain.

Lot / Code Info

ERGO++ Release 1.6.3 and 1.6.3.1

Root Cause

Software design

Action Taken

Computerized Medical Systems, Inc. sent an "USER NOTIFICATION" dated December 11, 2009, to all affected customers. The notice identified the product, problem and suggested a workaround to minimize the risk. The notice stated that this workaround does NOT substitute the need for regular Quality Assurance. Customers are instructed to review page 39 of user manual for further explanation of warnings. Customers were asked to complete the Notice for Customer section and return to 3D Line Medical System. This issue has been resolved in Release 1.7.0 and above. No follow-up notice will be distributed to customers. For further information please contact: CMS Software (314) 812-4460 (St Louis Office)