Monaco Radiation Treatment Planning System, Version 3.10.00. Planning of radiation therapy.

This recall has been terminated (originally issued April 3, 2012).

Company
Computerized Medical Systems Inc
Recall Initiated
March 5, 2012
Posted
April 3, 2012
Terminated
April 1, 2014
Recall Number
Z-1365-2012
Quantity
44
Firm Location
Maryland Heights, MO
Official Source
View on FDA website ↗

Reason for Recall

An unnecessary dose refresh can occur in certain situations which causes inconsistencies in the Beam Doses and the Total Dose.

Distribution

Worldwide Distribution -- USA, including the states of OH, IN, WI, CA, FL, IL, LA, NJ, MA, NC, and MO and the countries of Canada, Australia, Belgium, China, France, Germany, Japan, The Netherlands, New Zealand, Portugal, Switzerland, Taiwan, and the United Kingdom.

Lot / Code Info

Version 3.10.00

Root Cause

Software design

Action Taken

An "IMPORTANT SAFETY NOTICE" was issued between 3/5-9/12 via e-mail, FAX, or traditional mail. The Notice described the correct functioning of the system, the problem, when and why the issue occurs, the clinical impact, and the workaround. Patch 3.10.01 is now available. Customers can download their patch at the link provided in the Notice. The Notice also provided a list of offices where customers can contact if they have questions or further support is needed.