Hitachi Oasis Magnetic Resonance Scanners Software V1.0C The intended use: The MR system produces transverse, coronal, sagittal, oblique, and curved cross-sectional images that display the internal structure of the head, body, or extremities. The images produced by the MR system reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance.

This recall has been terminated (originally issued May 4, 2010).

Company
Hitachi Medical Systems America Inc
Recall Initiated
September 1, 2009
Posted
May 4, 2010
Terminated
February 24, 2012
Recall Number
Z-1534-2010
Quantity
43
Firm Location
Twinsburg, OH
Official Source
View on FDA website ↗

Reason for Recall

The device has a software error that could potentially lead to a misdiagnosis.

Distribution

Nationwide distribution: AZ, CA, CT, DE, FL, GA, IA, IL, IN, LA, MD, MI, MO, NC, NY, OH, OK, OR, PA, RI, SC, TX, VA, WA, WI, and WY.

Lot / Code Info

Model #'s: M001 - M043

Root Cause

Software design (manufacturing process)

Action Taken

Sites have already been corrected. Software updated from V1.0C to V2.0B