Hitachi CXR4 Computed Tomography Scanner. This product is an x-ray imaging device that produces cross-sectional images of he body at different angles. The system reconstructs, processes, displays, and stores the collected images. The device output can provide an aid to diagnosis when used by a qualified physician and is intended for general purpose CT applications.

This recall has been terminated (originally issued January 8, 2009).

Company
Hitachi Medical Systems America Inc
Recall Initiated
November 5, 2008
Posted
January 8, 2009
Terminated
September 21, 2012
Recall Number
Z-0608-2009
Quantity
42
Firm Location
Twinsburg, OH
Official Source
View on FDA website ↗

Reason for Recall

Software error: The device has a software error that can result in a condition where the displayed image orientation does not match the display orientation markers (right-left reversal).

Distribution

Nationwide Distribution --- including states of AZ, CA, DE, FL, ID, IL, IN, KS, KY, LA, MA, MO, NJ, NY, OH, OK, PA, PR, SC, TX, VA, and WI.

Lot / Code Info

Serial numbers: CXR46202 through CXR46248 excluding CXR46207, CXR46208, CXR46215, and CXR46222.

Root Cause

Software design

Action Taken

On 11/5/2008 Hitachi Medical Systems sent recall notification letters entitled "URGENT - DEVICE RECALL CORRECTION" to its customers informing them of the problem with the software. They stated that a Hitachi service representative will contact each customer to schedule a software upgrade. For further information, please contact Hitachi Medical Systems America Inc. by telephone at 330-425-1313.