SoftPath GUI version 4.2.3.4 2008-06-30 (Manuf Date). SoftPath GUI version 4.2.4.1 2008-09-30 (Manuf Date). SoftPath GUI version 4.3.0.3, 2008-07-23 (Manuf Date). SoftPath GUI version 4.3.0.4 2008-09-16 (Manuf Date). SCC Soft Computer, 5400 Tech Data Drive, Clearwater, FL 33760.

This recall has been terminated (originally issued October 20, 2010).

Company
Scc Soft Computer
Recall Initiated
January 26, 2009
Posted
October 20, 2010
Terminated
November 3, 2010
Recall Number
Z-0102-2011
Quantity
52
Firm Location
Clearwater, FL
Official Source
View on FDA website ↗

Reason for Recall

For SoftPath GUI versions 4.2.3.4, 4.2.4.1, 403.0.3, and 4.3.0.4., the Previous Patient History and some QA Management reports may have missing or incorrect diagnosis text. However, the patient final report, HIS resulting and Query display are all correct. All affected clients have been provided corrected software..

Distribution

Worldwide Distribution: Throughout USA and Canada.

Lot / Code Info

SoftPath GUI versions 4.2.3.4, 4.2.4.1, 4.3.0.3, 4.3.0.4.

Root Cause

Software design

Action Taken

SCC Soft Computer issued a Risk-to-Health notification with Client Notification Verbiage to affected clients using SCC Soft Computer's proprietary Task Management System on January 26, 2009. SCC issued a software upgrade with corrective patch electronically to all clients. Client was asked to acknowledge receipt of the task and grant permission to load the required software. SCC will contact each client to coordinate the running of the utility and the loading and validation of the correct patch.