RecallDepth

Anti-HBs Reagent Kit, list 7C18-25, 100 tests; Abbott Diagnostics, a Division of Abbott Laboratories, Abbott Park, IL 60064

This recall has been terminated (originally issued May 9, 2003).

Company
Abbott Laboratories Hpd/Add
Recall Initiated
April 7, 2003
Posted
May 9, 2003
Terminated
August 19, 2003
Recall Number
Z-0801-03
Quantity
289 kits
Firm Location
Abbott Park, IL
Official Source
View on FDA website ↗

Reason for Recall

Particulate matter in the microparticle component.

Distribution

Canada, Germany and Japan

Lot / Code Info

List 7C18-25, lot 95305M200

Root Cause

Other

Action Taken

Abbott affiliates in Canada, Germany and Japan were faxed copies of the recall letter on 4/7/03 for translation where needed and dissemination to their accounts receivin the suspect lot. The end users were advised of the mold-microparticle aggregation and instructed to discontinue use and destroy any of the lot remaining.