Free T4 Control Pack, list 9C03-10; the kit consists of 3 bottles of controls, L - 0.7, M - 1.2 and H - 3.0; Abbott Laboratories, Abbott Park, IL 60064

This recall has been terminated (originally issued August 13, 2003).

Company
Abbott Laboratories Hpd/Add
Recall Initiated
May 30, 2003
Posted
August 13, 2003
Terminated
September 16, 2003
Recall Number
Z-1130-03
Quantity
3247 kits
Firm Location
Abbott Park, IL
Official Source
View on FDA website ↗

Reason for Recall

The Free T4 Controls, may yield control values that are high and outside of the package insert ranges.

Distribution

Nationwide and internationally to Abbott subsidiaries in Ecuador, Uruguay, Argentina, Brazil, El Salvador, Canada, Germany, Japan, Singapore, Hong Kong, Thailand, Korea, Australia, Costa Rica, Grenada, St. Lucia, Bermuda, Antigua, Panama, Guatemala, Mexico, Colombia, Jamaica, Paraguay, Peru, Chile, Dominican Republic, United Kingdom, New Zealand, Trinidad & Tabago, St. Vincent, Barbados, Curacao, St. Kitts, Anguilla, Turks & Caicos Islands, Cayman Islands, Germany and Taiwan.

Lot / Code Info

List 9C03-10, lot numbers 95881Q100 and 95881Q101

Root Cause

Other

Action Taken

The Free T4 Control Pack, list 9C03-10, and the IMx Free T4 Reagent Kit, list 2222-20, were recalled by letter dated 5/30/03. The accounts were informed of the out of range control values, and the potential for depressed patient values with the IMx kit, and were requested to discontinue use and destroy the affected lots of product. Abbott sent a Device Correction letter dated 6/4/03 to those accounts receiving the Architect Free T4 Calibrator kits, informing them of the problems experienced with the The Free T4 Control Pack, list 9C03-10, and the IMx Free T4 Reagent Kit, list 2222-20, made from the same raw material. The accounts were informed that the Architect Calibrator was performing as expected, but as a precaution recommended conducting a daily QC with non-Abbott Controls and to report back to Abbott any out of range non-Abbott control results.