AxSYM AFP Reagent Pack, list 7A48-22 (100 test kit), 7A48-21(French label); and AxSYM Cancer Assay Disk, list 3D50-02 (Version 2.0), 3D50-03 (Version 3.0), and 3D50-04 (Version 4.0)

This recall has been terminated (originally issued July 24, 2003).

Company
Abbott Laboratories Hpd/Add
Recall Initiated
June 12, 2003
Posted
July 24, 2003
Terminated
November 21, 2003
Recall Number
Z-1047-03
Quantity
23,000 kits
Firm Location
Abbott Park, IL
Official Source
View on FDA website ↗

Reason for Recall

AFP results obtained when using the 1:101 automated dilution protocol showed an overestimation of AFP of up to 30.5%

Distribution

Nationwide and internationally to Australia, Canada, Costa Rica, Ecuador, Japan, Panama, San Salvador, Thailand, Barbados, Chili, Curacao, Honduras, Korea, Peru, Singapore, United Kingdom, Brazil, Columbia, Dominican Republic, Hong Kong, Pakistan, St. Vincent, Uruguay

Lot / Code Info

All lots of AFP list 7A48-22 and 7A48-21 used with all lots of AxSYM Cancer Assay Disk list 3D50-02, 3D50-03 and 3D50-04 with AFP Assay File version 2.00.200, when used with AxSYM SYstem Sotware Versions 3.00 and higher.

Root Cause

Other

Action Taken

All AxSYM AFP customers were informed not to use the 1:101 automated dilution protocol via Device Correction letter dated 6/12/03. If necessary, manually dilute samples per the AxSYM AFP package insert.