RecallDepth

14F x 24CM SLX Double Lumen Full Tray Hemodialysis catheter insertion kits, Item No. MC061424 Product Usage: An aid for introducing a hemodialysis catheter

Company
Medical Components, Inc Dba Medcomp
Recall Initiated
March 13, 2019
Terminated
March 5, 2024
Recall Number
Z-2216-2019
Quantity
6830
Firm Location
Harleysville, PA

Reason for Recall

Tray label indicates the kits are packaged with a 15F "valved" peelable introducer, but the kits contain a 15F "non-valved" peelable introducer.

Distribution

US Nationwide Distribution CA FL GA IN KY LA MA MI MO MS NJ NY OH OK PA SC TN TX VA WV.

Lot / Code Info

MCAL430 MCAT000 MCBM770 MCBS380 MCCC840 MCCM930 MCCS360 MCCZ450 MCDC580 MCDN120 MCDR420 MCFB350 MCFG980 MCFM410 MCFP230 MCFT430 MCFZ290 MKAF350 MKAG490 MKAM210 MKAR120 MKBA200 MKBL090 MKBL220 MKBV580 MMAB710 MMAM070 MMAQ950

Root Cause

Labeling Change Control

Action Taken

On March 13, 2019, the firm notified customers via Urgent Medical Device Recall letter. Customers were informed that the kits contain a 15F "non-valved" peelable introducer despite the label indicating that the kit is packaged with a "valved" introducer. Customers were asked to do the following: 1. Immediately examine your inventory and quarantine product subject to recall. 2. If you have further distributed this product, please identify your customers and immediately notify them of the recall. 3. Contact your Medical Components, Inc. customer service representative for a Returned Goods Authorization (RGA) number if you have product to return. The phone number to call is 215-256-4201. 4.Complete and return the Medical Device Recall Return Response Form. The form can be returned by fax (215-256-9191) or email (recalls@medcompnet.com). Contact Information for questions: Monday through Friday, 8:00 AM to 4:00 PM, Eastern Standard Time. Phone: 215-256-4201; Email: recalls@medcompnet.com

More recalls by Medical Components, Inc Dba Medcomp

View all recalls by this company →