Terumo Advanced Perfusion System 1 Electronic Patient Gas System Catalog #: 801188
- Recall Initiated
- June 8, 2021
- Terminated
- June 28, 2022
- Recall Number
- Z-2350-2021
- Quantity
- 1 unit OUS
- Firm Location
- Ann Arbor, MI
Reason for Recall
intermittent failure causes a blender initialization fault upon start-up, or during an EPGS calibration attempt following the 15-minute warm-up period. The control sliders on the Central Control Monitor (CCM) are disabled and the gas system can only be controlled by the local control knobs
Distribution
International distribution in the country of Panama.
Lot / Code Info
S/N: 01751 UDI: 00886799000588
Root Cause
Nonconforming Material/Component
Action Taken
Terumo notified the regional affiliate to address the single EPGS device in Panama via email on 6/8/21. Letter states reason for recall, health risk and action to take: Terumo CVS will execute a field correction to replace the one (1) affected EPGS device. A Terumo Representative will contact you to schedule the field correction activities. 1. Review this Medical Device Correction and assure the user has received notice of this issue. 2. A Terumo representative will contact the user to schedule the field correction activity. Note: Terumo CVS recommends that users continue using the affected Terumo System 1 while waiting for this correction. We encourage you to contact your local Terumo representative with any questions or concerns.