Terumo CDI OneView Monitoring System Hematocrit / Oxygen Saturation Probe, Catalog Number REF CD1754

Company
Terumo Cardiovascular Systems Corporation
Recall Initiated
March 3, 2025
Posted
March 31, 2025
Recall Number
Z-1492-2025
Quantity
37 units
Firm Location
Ann Arbor, MI

Reason for Recall

During some clinical cases, it has been observed that the SO2 value may periodically appear as dashes on the Touchscreen Display indicating unavailable data. In these cases, a yellow low-limit alarm will also be triggered to the user, due to the dashed-out value.

Distribution

US: IL, TX, NY, MI, FL, NV, and Australia, New Zealand

Lot / Code Info

UDI/DI 00886799001882, Serial Number range: H0050095-H0050110, H0050112-H0050116, H0050118-H0050124, H0050126, H0050131-H0050134, H0050136, H0050140-H0050142

Root Cause

Software design

Action Taken

Terumo Cardiovascular issued an URGENT MEDICAL DEVICE REMOVAL notice to its consignees on 03/03/2025 via email. The notice explained the reason for the removal, potential hazard, and requested the following: 1. Review this Medical Device Correction and assure that all users have received notice of this issue. 2. Confirm receipt of this communication by completing and returning the attached Customer Response Form as indicated on the form. 3. Terumo CVS will contact users to discuss removal and correction activities. Note: Terumo CVS recommends that users continue using CDI OneView System H/S Probes while waiting for this removal