The Sarns Temperature Control and Monitor unit (TCM) is a source of temperature-controlled water for blood heat exchangers used in an extracorporeal circuit and for blankets to externally heat or cool the patient. The TCM with options will also supply water for cardioplegia, freeze water for an ice supply, monitor temperatures in the patient and extracorporeal circuit, and allow gradient rewarming relative to a venous blood temperature. Device Name / Model Number: TCM II TUV, 115V (P/N 4415), TCM II TUV, 220V (P/N 4416), Catalog Number: 4416, 164940

Company
Terumo Cardiovascular Systems Corporation
Recall Initiated
April 16, 2021
Terminated
November 29, 2023
Recall Number
Z-1790-2021
Quantity
1176 devices
Firm Location
Ann Arbor, MI

Reason for Recall

Terumo CVS has been unable to validate a cleaning protocol to satisfy current regulatory concerns and expectations. As a result, an updated cleaning protocol will not be developed by Terumo CVS and it has been determined that the best course of action is for users to discontinue use of and dispose of HX2, TCM I and TCM II devices.

Distribution

Domestic: Foreign: Australia, Belgium, Canada, Chile, China, Colombia, Dominican Republic, England, France, Germany, Greece, Hong Kong, India, Indonesia, Iran, Israel, Italy, Japan, Korea, Malaysia, Mexico, Netherlands, New Zealand, Norway, Philippines, Russia, Saudi Arabia, Singapore, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, UNITED ARAB EMIRATES (UAE), Vietnam

Lot / Code Info

All lot numbers distributed from 05/02/1985 thru 06/10/2015

Root Cause

Device Design

Action Taken

On 04/30/2021, Terumo issued an Urgent Medical Device Removal notice to customer via letter notifying users to discontinue the use of and dispose of HX2, TCM I and TCM II devices. Customers were instructed to confirm receipt of this communication by completing and returning the attached Customer Response Form as indicated on the form. For questions contact Terumo CVS Customer Service: 1-800-521-2818.