OEC 9800 Image Intensified fluoroscopic x-ray system (includes OEC 9800, OEC FluoroTrak 9800 Plus, OEC 9800 Plus, and OEC 9800MD Motorized C-arm System), GE Healthcare, Surgery, Salt Lake City, UT.

This recall has been terminated (originally issued September 1, 2008).

Company
Oec Medical Systems, Inc
Recall Initiated
April 15, 2008
Posted
September 1, 2008
Terminated
December 21, 2011
Recall Number
Z-1705-2008
Quantity
78 units
Firm Location
Salt Lake City, UT
Official Source
View on FDA website ↗

Reason for Recall

Use of existing four-pedal footswitch on a different machine may cause various operational errors.

Distribution

Nationwide, including one VA facility in FL. No military or foreign distribution.

Lot / Code Info

Serial numbers: 82-0886, 82-1387, 82-1471, 82-1827, 82-1837, 82-2088, 82-2091-C, 82-2219, 82-2265, 82-2298, 82-2308, 82-2330, 82-2342, 82-2418, 82-2633, 82-2764, 82-2902, 82-2919, 82-2982, 82-3012-N, 82-3033, 82-3062, 82-3147, 82-3185, 82-3311, 82-3313, 82-3314, 82-3411, 82-3428-E, 82-3509, 82-3510, 82-3576, 82-3739, 82-3763, 82-3765, 82-3830, 82-3880, 82-7106, 82-7179-MH, 89-0014, 89-0017, 89-0938, 89-1128, 89-1451, 89-2212, 89-2402, 89-2773, 89-2989, 89-3078, 89-3083, 89-3164, 89-3301, 89-3356, 89-3487, 8S-0271, 8S-0488, 8S-0974, 8S-1460, 8S-1487, 8S-1508-N, 8S-1582, 8S-1932-N, 8S-2425, 8S-2507, 8S-3054, 8S-3236, 8S-3242, 8S-7141-MH, 8S-7199-MH.

Root Cause

Device Design

Action Taken

Consignees were notified by letter on 03/21/2007 and advised that GE would be replacing the 4-pedal switches at some time in the future. Replacement began on 04/15/2008. Contact GE Healthcare at 1-800-874-7378 for assistance.