GE Healthcare OEC 9900 mobile fluoroscopic x-ray system

This recall has been terminated (originally issued September 29, 2008).

Company
Oec Medical Systems, Inc
Recall Initiated
December 14, 2007
Posted
September 29, 2008
Terminated
April 13, 2012
Recall Number
Z-0368-2008
Quantity
710
Firm Location
Salt Lake City, UT
Official Source
View on FDA website ↗

Reason for Recall

Possible overexposure: Failure to apply the entrance exposure rate (EER) tube current limit calibration to the automatic exposure rate (AER) control system when the anatomical profile mode is changed from the default selection to another selection. Manual technique modes are not affected.

Distribution

Worldwide distribution: USA, Finland, Netherlands, Germany, Switzerland, France, United Kingdom, Australia, Italy, Ireland, Spain, Sweden, Belgium, New Zealand, China, Singapore, Poland, Norway, and India.

Lot / Code Info

All codes

Root Cause

Software design

Action Taken

GE OEC sent a letter to customers December 15, 2007, and scheduled field representative visits for corrective software solution installation.