Accuracy Cyberknife Robotic Radiosurgery System, medical charged particle radiation therapy device. Model number 020700 (axum/standard treatment couch) and 021756 hand controller. The device is indicated for treatment planning and image-guided stereotactic radiosurgery and precision radiotherapy of lesions, tumors and conditions anywhere in the body when radiation treatment is indicated.
This recall has been terminated (originally issued September 16, 2008).
- Company
- Accuray Inc
- Recall Initiated
- May 1, 2008
- Posted
- September 16, 2008
- Terminated
- November 30, 2010
- Recall Number
- Z-1862-2008
- Quantity
- 105 units
- Firm Location
- Sunnyvale, CA
- Official Source
- View on FDA website ↗
Reason for Recall
Couch may move unexpectedly, which may result in patient impacting the linear accelerator.
Distribution
Product was distributed throughout the US and to China, Taiwan, Korea, Italy, Malaysia, Germany, Netherlands, Vietnam, Greece, Spain, and Puerto Rico.
Lot / Code Info
All models.
Root Cause
Software design
Action Taken
All affected sites were notified via customer advisory notice. The Standard Treatment couch firmware will be updated and all affected sites will be retrofitted. Contact Customer Service at 1-877-668-8667 (USA) or 1-408-716-4700 (non-USA) or customersupport@accuray.com for additional information.
More recalls by Accuray Inc
Radiation therapy device. Medical charged particle radiation therapy system with software version....
Feb 2, 2011
CyberKnife Treatment Planning System, a subsystem of the CyberKnife Robotic Radiosurgery System, wit...
Feb 1, 2011
CyberKnife System Robotic Radiosurgery System, (Robocouch Patient positioning System) Synchrony Resp...
Jan 18, 2011