Accuracy Cyberknife Robotic Radiosurgery System, medical charged particle radiation therapy device. Model number 020700 (axum/standard treatment couch) and 021756 hand controller. The device is indicated for treatment planning and image-guided stereotactic radiosurgery and precision radiotherapy of lesions, tumors and conditions anywhere in the body when radiation treatment is indicated.

This recall has been terminated (originally issued September 16, 2008).

Company
Accuray Inc
Recall Initiated
May 1, 2008
Posted
September 16, 2008
Terminated
November 30, 2010
Recall Number
Z-1862-2008
Quantity
105 units
Firm Location
Sunnyvale, CA
Official Source
View on FDA website ↗

Reason for Recall

Couch may move unexpectedly, which may result in patient impacting the linear accelerator.

Distribution

Product was distributed throughout the US and to China, Taiwan, Korea, Italy, Malaysia, Germany, Netherlands, Vietnam, Greece, Spain, and Puerto Rico.

Lot / Code Info

All models.

Root Cause

Software design

Action Taken

All affected sites were notified via customer advisory notice. The Standard Treatment couch firmware will be updated and all affected sites will be retrofitted. Contact Customer Service at 1-877-668-8667 (USA) or 1-408-716-4700 (non-USA) or customersupport@accuray.com for additional information.