RoboCouch Patient Support System, model number 025007, a subsystem of the Cyberknife Robotic Radiosurgery System, manufactured by Accuray Incorporated, Sunnyvale, CA. Intended for use in the support and positioning of a patient during Radiosurgery and radiotherapy procedures and other medical procedures when precise positioning is required.

This recall has been terminated (originally issued January 11, 2011).

Company
Accuray Inc
Recall Initiated
July 21, 2009
Posted
January 11, 2011
Terminated
January 12, 2011
Recall Number
Z-0853-2011
Quantity
26
Firm Location
Sunnyvale, CA
Official Source
View on FDA website ↗

Reason for Recall

Mechanical Failure-- Patient Support System's table top may unexpectedly descend several inches.

Distribution

Worldwide Distribution -- USA, including states of CA, FL, VA, AL, CT, MA, NC, TN, PA, RI, and NE and countries of Turkey, France, UK, Saudi Arabia, and Switzerland.

Lot / Code Info

RoboCouch model 025007

Root Cause

Device Design

Action Taken

Consignees were notified verbally about the issue when an inspection was scheduled.