CyberKnife System Robotic Radiosurgery System, (Standard Treatment Couch) Synchrony Respiratory Tracking System, a Radiation Therapy Device, Model Numbers: 026429, 028185, Product is manufactured and distributed by Accuray, Inc., Sunnyvale, CA A Radiation therapy device.
This recall has been terminated (originally issued January 18, 2011).
- Company
- Accuray Inc
- Recall Initiated
- March 16, 2009
- Posted
- January 18, 2011
- Terminated
- January 19, 2011
- Recall Number
- Z-0908-2011
- Firm Location
- Sunnyvale, CA
- Official Source
- View on FDA website ↗
Reason for Recall
The product has the potential for the extension coupling of the Synchrony extension mounting hardware to become loose over time. The assembly weighs approximately 17 pounds; an unexpected contact with users or patients may result in serious injury if the boom descends unexpectedly.
Distribution
Worldwide Distribution - USA (nationwide) and the countries of Turkey, Russia, France, Italy, the Netherlands, the UK, Spain, Greece, Switzerland, Germany, India, Taiwan, China, Hong Kong, Korea, Thailand, Malaysia and Japan
Lot / Code Info
All units of these models.
Root Cause
Device Design
Action Taken
Accuracy Inc., sent an URGENT ADVISORY NOTIFICATION letter, dated March 16, 2009, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to inspect their system to see if it was affected. Customers were instructed to contact Accuracy Customer Support at 1-877-668-8667 (USA) or 1-408-716-4700 or e-mail customersupport@accuracy.com to schedule service.