RecallDepth

Harleco May and Grunwald's Stain; 0.17% w/v; Methanol Solution UN1230; EMD Chemicals Inc. An Affilifate of Merck KGaA Damstadt, Germany; 480 South Democrat Road; Gibbstown, NJ 08027; 1 liter and 4 liter poly bottles

This recall has been terminated (originally issued November 3, 2007).

Company
Emd Chemicals Inc
Recall Initiated
July 30, 2007
Posted
November 3, 2007
Terminated
December 26, 2007
Recall Number
Z-0133-2008
Quantity
26 units (1 liter), 1 unit (4 liter)
Firm Location
Gibbstown, NJ
Official Source
View on FDA website ↗

Reason for Recall

Out of Spec -- The Batch (Lot) was incorrectly released after being rejected for out of specification of high azure and eosin absorbance at the end of the run and particulate matter at the end of the run.

Distribution

Product was released to 2 distributors; one in the US, one in Canada

Lot / Code Info

Item # 660/75 Lot 7161(1 liter bottle) Item 660/85, Lot 7161(4 liter bottle) Exp. Date: 31 Jan 09

Root Cause

Employee error

Action Taken

EMD Chemicals via fax and mailed an Important Product Recall letter dated 7/30/07. The letter notified end use consignee that after conducting an investigation it has determined that Harleco May and Grunwald's Stain; 0.17% w/v Item #660/75 % 660/85 has unacceptable levels of azure and eosin absorbance and undissolved dye matter. Users were requested to discontinue shipping and return product. Wholesalers were advised this is a Customer Level Product Recall and requested to sub recall product sold after 7/5/07.