RecallDepth

Anaerotest for microbiology Specification: Performance Test Anerobic: white within 4-6 hours; aerobic: blue with 20 minutes Merck KgaA 64271 Darmstadt, Germany 1.19034.0002 17.08.06 1 unit UPC 4 022536 667928

This recall has been terminated (originally issued February 17, 2007).

Company
Emd Chemicals Inc
Recall Initiated
August 17, 2006
Posted
February 17, 2007
Terminated
June 6, 2008
Recall Number
Z-0498-2007
Quantity
8 rolls
Firm Location
Gibbstown, NJ
Official Source
View on FDA website ↗

Reason for Recall

Lot # OC553145 of Anaerotest failed to remain in specification. QC indicator shows an anaerobic environment even if oxygen is present.

Distribution

Product was only sent to one US customer: Becton Dickinson, Cayey. Pureto Rico.

Lot / Code Info

Item number 1.19034.0002 Lot # OC553145

Root Cause

Other

Action Taken

A Recall Letter was issued by EMD Chemicals on 8/17/2006 to discontinue product usage and for product return.