Anaerotest;Microbiology , 50 test strips 1.15112 Store dry and tightly closed. Store at + 15 C to +25C Merck IVD 64271 Darmstadt, Germany
This recall has been terminated (originally issued February 17, 2007).
- Company
- Emd Chemicals Inc
- Recall Initiated
- August 17, 2006
- Posted
- February 17, 2007
- Terminated
- June 6, 2008
- Recall Number
- Z-0499-2007
- Quantity
- 800 packages of 50 strips each.
- Firm Location
- Gibbstown, NJ
- Official Source
- View on FDA website ↗
Reason for Recall
Lot # OC553145 of Anaerotest failed to remain in specification. QC indicator shows an anaerobic environment even if oxygen is present
Distribution
Product was only sent to one US customer: Becton Dickinson, Cayey. Pureto Rico.
Lot / Code Info
Item # 1.15112.0001 Lot # OC553145
Root Cause
Other
Action Taken
A Recall Letter was issued by EMD Chemicals on 8/17/2006 to discontinue product usage and for product return.
More recalls by Emd Chemicals Inc
Harleco Giemsa Stain 25 gram glass bottle, Item #228-12 Manufactured for EMD Chemicals Inc., Gibbsto...
Oct 12, 2011
Histochemical PAS (Periodic Acid Schiff) Reaction Set'
Class I, IVD;
Set of 4 bottles including Schi...
Sep 20, 2008
EMD Chemicals, Harleco Brand Ethanol Standard, 1.0 mg/ml;
1.0 mg Ethanol in 1 ml, 10 ampules; Item n...
Sep 16, 2008