RecallDepth

Anaerotest;Microbiology , 50 test strips 1.15112 Store dry and tightly closed. Store at + 15 C to +25C Merck IVD 64271 Darmstadt, Germany

This recall has been terminated (originally issued February 17, 2007).

Company
Emd Chemicals Inc
Recall Initiated
August 17, 2006
Posted
February 17, 2007
Terminated
June 6, 2008
Recall Number
Z-0499-2007
Quantity
800 packages of 50 strips each.
Firm Location
Gibbstown, NJ
Official Source
View on FDA website ↗

Reason for Recall

Lot # OC553145 of Anaerotest failed to remain in specification. QC indicator shows an anaerobic environment even if oxygen is present

Distribution

Product was only sent to one US customer: Becton Dickinson, Cayey. Pureto Rico.

Lot / Code Info

Item # 1.15112.0001 Lot # OC553145

Root Cause

Other

Action Taken

A Recall Letter was issued by EMD Chemicals on 8/17/2006 to discontinue product usage and for product return.