Maquet PowerLED 700. Product Usage; To provide illumination to the surgical area or the patient.
- Company
- Maquet Medical Systems, Usa
- Recall Initiated
- April 16, 2014
- Posted
- June 6, 2014
- Terminated
- April 6, 2016
- Recall Number
- Z-1730-2014
- Quantity
- United States 1169 units
- Firm Location
- Wayne, NJ
Reason for Recall
Maquet has identified an issue with the Maquet PowerLED 700 (certain serial numbers). One or several plastic fixation tabs can break, which can cause the ambient light module to partially detach.
Distribution
US Nationwide Distribution.
Lot / Code Info
Part number (begins with ALM) and associated serial numbers: ALM568300999 13, 14, 16, 19, 20, 21, 22, 23, 24, 25, 29, 30, 31, 32, 34, 35, 36, 41, 50 ALM568301999 13, 14, 16, 17, 20, 23, 24, 25, 26, 52, 54, 55, 56, 57, 60, 63, 64, 65, 66, 70, 71, 72, 74 ALM568302999 11, 43, 44, 46, 49, 50, 51, 54, 57, 58, 59, 6, 61, 63, 64, 66, 67, 71, 72, 74, 77, 78, 80, 81 ALM568303999 8, 9, 32, 52, 53, 58, 60, 61, 62, 65, 66, 69, 70, 71, 73, 74, 75, 77, 78, 79, 80, 81, 82, 85, 86, 87, 89, 91, 92, 94, 95, 96, 97, 99, 100, 105, 106, 108, 111, 118, 120, 121, 122, 159 ALM568320901 12012, 40015, 40017, 40018, 40019, 40020, 40021, 50010, 50011, 50012, 50013, 50014, 50015, 50016, 50017, 50018, 50019, 50020, 50021, 50022, 50023, 50024, 50025, 50026, 50027, 50028, 50030, 50031, 50032, 50033, 50034, 50035, 50037, 50038, 50039, 50040, 50041, 50042, 50043, 50047, 50048, 50050, 50051, 50052, 50053, 50054, 50055, 50056, 50058, 50059, 50060, 50061, 50064, 50066, 50067, 50068, 50069, 50070, 50071, 50072, 70033, 70034, 70035 ALM568320905 40003, 40004, 40005, 40006, 40007, 40008, 40009, 40010, 40011, 40012, 40013, 40014, 40015, 40016, 50003, 50004, 50005, 50006, 50008, 50009, 50010, 50011, 50012, 50013, 50014, 50015, 50016, 50017, 50018, 50019, 50020, 50021, 60003, 60004, 70001, 70002, 70003, 70004, 70005, 70006, 70007, 70009, 70010, 70011, 70012, 70013, 70014, 70015, 70016, 70017, 70018, 70019, 70020, 70021, 70022, 70024, 70025, 70027, 70028, 70029, 70030, 70031, 70032, 70033, 70034, 70035, 70036, 70037, 70038, 70039, 70040, 70041, 70042, 70043, 70044, 70045, 70046, 70047, 70048, 70049, 70050, 70051, 70052, 70053, 70054, 70055, 70056, 70057, 70058, 70059, 70060, 70061, 70063, 70064, 70066, 70068, 70069 ALM568370900 20017, 20018, 20019, 20020, 20021, 20022, 20023, 20024, 20025, 20026, 20027, 20028, 20029, 20030, 20039, 20040, 20041, 20042, 20043, 20044, 20045, 20046, 20047, 20048, 20049, 20050, 20051, 20052, 20053, 20054, 30038, 30039, 30053, 30054, 40006, 40010, 40012, 40013, 40014, 40015, 40016, 40018, 40019, 40020, 40027, 40028, 40029, 40030, 40031, 40032, 40033, 40034, 50000, 50001, 50004, 50006, 60007, 60008, 60009, 60010, 60012, 60031, 60041, 60042, 60043, 60044, 60046, 60047, 60048 ALM568370901 20024, 20025, 20026, 20027, 20028, 20029, 20030, 20031, 20032, 20033, 20034, 20038, 20039, 20040, 20041, 20042, 20045, 20053, 20054, 20055, 20056, 20057, 20058, 20059, 20060, 20061, 20062, 20064, 20065, 20066, 20067, 20068, 20069, 20070, 32042, 40028, 40033, 40034, 40035, 40036, 40037, 40038, 40045, 50002, 50006, 50008, 50009, 50024, 50035, 50036, 50037, 50039, 50040, 50041, 50042, 50043, 50044, 50045, 50046, 50047, 50048, 60030 ALM568370902 20011, 20015, 20016, 20017, 20018, 20019, 20020, 20021, 20022, 20023, 20024, 20025, 20026, 20027, 20028, 20029, 20030, 20031, 20032, 20033, 20034, 20035, 20036, 20037, 20038, 20039, 20040, 20041, 20042, 20048, 20049, 20050, 20051, 20057, 20058, 20059, 20060, 20061, 20062, 20063, 20064, 20067, 20068, 20069, 20070, 20071, 20072, 20073, 20074, 20075, 20076, 20077, 20078, 20079, 20080, 20083, 20086, 20087, 20089, 20090, 20091, 20094, 20095, 20096, 20098, 20104, 20105, 20106, 20107, 20108, 20109, 20110, 20111, 20112, 20113, 20114, 20115, 20116, 20117, 20118, 20119, 20120, 20121, 20122, 20123, 20124, 20125, 20127, 20132, 20133, 20135, 20136, 20137, 20139, 20140, 20141, 20144, 20151, 20152, 30000, 30051, 30053, 30054, 30062, 30063, 30065, 30066, 30067, 30068, 30078, 30081, 30090, 30091, 30092, 30099, 30107, 30127, 30128, 30130, 30131, 30134, 32164, 40000, 40006, 40007, 40009, 40011, 40012, 40013, 40014, 40015, 40017, 40018, 40019, 40020, 40021, 40023, 40039, 40040, 40041, 40042, 40043, 40044, 40045, 40046, 40047, 40048, 40050, 40051, 40052, 40053, 40054, 40055, 40056, 40057, 40058, 40060, 40064, 40065, 40067, 40078, 40081, 40082, 40083, 40084, 60030, 60031, 60036 ALM568370903 20022, 20023, 20024, 20025, 20026, 20027, 20028, 20029, 20030, 20032, 20033, 20034, 20035, 20036, 20037, 20038, 20039, 20041, 20043, 20044, 20045, 20046, 20047, 20048, 20049, 20050, 20051, 20052, 20053, 20055, 20056, 20057, 20058, 20059, 20066, 20067, 20072, 20073, 20074, 20075, 20076, 20077, 20079, 20080, 20081, 20082, 20084, 20087, 20088, 20089, 20090, 20091, 20092, 20093, 20097, 20098, 20099, 30072, 40025, 50000, 50001, 50002, 50003, 50004, 50006, 50028, 50051, 50052, 50054, 50055, 50057, 50058, 50059, 50060, 60024, 60025, 60026, 60028 ALM568370904 20011, 20013, 20014, 20015, 20016, 20017, 20018, 20019, 20020, 20023, 20024, 20025, 20026, 20027, 20032, 20033, 20034, 20035, 20036, 20037, 20038, 30000, 30001, 30002, 30003, 30004, 30012, 30013, 30014, 30015, 30018, 30019, 30020, 30025, 30026, 30027, 30028, 30029, 30030, 30031, 30032, 30033, 30034, 30035, 30036, 30037, 30038, 30039, 30040, 30041, 30042, 30043, 40011, 40012, 40013, 40014, 40015, 40016, 40017, 40022, 40023, 40024, 40031, 40032, 40033, 40034, 40035, 40036, 50004, 50010, 50011, 50012, 50013, 50014, 50015, 50016, 50017, 50019 ALM568370905 20010, 20016, 20017, 20018, 20019, 20022, 20023, 30027, 40012, 40022, 40023, 40024, 40025, 40027, 40028, 40029, 40031, 40032, 40033, 40037, 40040, 40041, 40047, 40049, 40050, 50005, 50006, 50012, 50015, 50018, 50021, 60007, 60023, 60027, 60030, 60032, 60033
Root Cause
Other
Action Taken
Maquet issued "Urgent-Medical Device Field Correction" notices, dated April 16, 2014 to affected customers via Federal Express. The notice identified the product, related issue and the action needed to be taken by the customer. The corrective action will include inspection of the fixation tabs of the Ambient Light Module adn replacement of the Ambient Light Module if tabs are found to be defective. The correction will be performed by an authorized MAQUET Service Technician. MAQUET Service Team representatives will schedule on-site service with customers. If customers have any questions or require additional information, the notice instructs them to contact their local MAQUET representative, or MAQUET Customer Service at 1-888-627-8383 (option 3), Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. EST.