CARDIOHELP-i System. Blood oxygenation and carbon dioxide removal system.

Company
Maquet Medical Systems, Usa
Recall Initiated
April 15, 2024
Posted
May 16, 2024
Recall Number
Z-1831-2024
Quantity
148 units (US: 5; OUS: 143)
Firm Location
Wayne, NJ

Reason for Recall

The firm identified that the measured patient leakage current in some devices significantly exceeded expected levels.

Distribution

US Nationwide distribution in the states of NJ & TX.

Lot / Code Info

Part No. 70104.8012 and 70104.8012R; UDI-DI: 04037691658384; Serial no. 90410021, 90410029, 90410165, 90410254, 90410255, 90410024, 90410025, 90410026, 90410028, 90410030, 90410031, 90410034, 90410035, 90410036, 90410043, 90410044, 90410045, 90410046, 90410047, 90410048, 90410049, 90410050, 90410051, 90410052, 90410053, 90410054, 90410055, 90410056, 90410057, 90410058, 90410059, 90410060, 90410061, 90410062, 90410063, 90410064, 90410065, 90410067, 90410068, 90410069, 90410070, 90410071, 90410072, 90410073, 90410074, 90410075, 90410076, 90410077, 90410078, 90410079, 90410080, 90410081, 90410082, 90410084, 90410085, 90410086, 90410087, 90410088, 90410089, 90410091, 90410093, 90410094, 90410095, 90410096, 90410097, 90410098, 90410099, 90410100, 90410101, 90410102, 90410103, 90410104, 90410105, 90410106, 90410107, 90410108, 90410109, 90410110, 90410111, 90410112, 90410113, 90410114, 90410115, 90410116, 90410117, 90410118, 90410119, 90410120, 90410121, 90410122, 90410123, 90410124, 90410125, 90410126, 90410127, 90410130, 90410131, 90410132, 90410133, 90410134, 90410135, 90410136, 90410137, 90410138, 90410139, 90410140, 90410141, 90410142, 90410143, 90410144, 90410146, 90410147, 90410148, 90410149, 90410151, 90410152, 90410153, 90410154, 90410155, 90410156, 90410157, 90410158, 90410159, 90410160, 90410161, 90410162, 90410163, 90410166, 90410167, 90410168, 90410169, 90410170, 90410172, 90410173, 90410174, 90410175, 90410176, 90410177, 90410178, 90410179, 90410180, 90410181, 90410182, 90410183, 90410184, 90410185, 90410186, 90410187.

Root Cause

Under Investigation by firm

Action Taken

Consignees received an "URGENT MEDICAL DEVICE - Removal" email, dated 4/15/2024, notifying them of this recall event. Consignees are asked to examine their inventory to identify if they possess any affected devices and all consignees are asked to return the provided Response Form via email to identify where affected devices are located. Consignees that report having any affected devices will be contacted by Getinge to arrange for return of the affected device to upgrade the system's sensor panel and perform testing to ensure that leakage current is within specifications. Technical Support can also be contacted at 888-943-8872 (option 4, option 2, option 1) Monday through Friday between 8:00 AM and 6:00 PM EST to make service arrangements. Consignees are asked to not use affected devices for any additional patients until it has been cleared by the MCP/Getinge system check, repaired, or replaced. The provided recall notice is to be forwarded to users of the CARDIOHELP-i system and if the device was further distributed.

More recalls by Maquet Medical Systems, Usa

View all recalls by this company →