Model No. BEQ-HLS 5050 USA, HLS Set Advanced 5.0, Product Code/Part Number 70106.9077. Used with the CARDIOHELP heart-lung machine.
- Company
- Maquet Medical Systems, Usa
- Recall Initiated
- January 16, 2024
- Posted
- February 13, 2024
- Recall Number
- Z-1076-2024
- Quantity
- 16520 worldwide; 6657 US
- Firm Location
- Wayne, NJ
Reason for Recall
The Emergency Priming Line, a component of the HLS Set, may have a compromised sterile barrier due to nonconformities at the bag seal. This issue may result in inflammation, sepsis, or infection.
Distribution
Worldwide - US Nationwide distribution.
Lot / Code Info
UDI-DI 04058863076355 Lots 3000301222, 3000319782, 3000327602 ***Added 3/21/24*** 3000333098, 3000352360, 3000357174
Root Cause
Nonconforming Material/Component
Action Taken
On January 16, 2024, the recalling firm began notifying affected customers via URGENT MEDICAL DEVICE CORRECTION letters. Customers were instructed to dispose of any Emergency Priming Line from a previously opened, affected HLS set. Customers should only use the Emergency Priming Line from a newly opened HLS Set. Device users should always have an HLS Set ready as a backup during therapy. The remaining components of the backup HLS Set must be disposed of after extraction of the Emergency Priming Line. Please reach out to Getinge Customer Support for the issuance of a replacement HLS Set at no cost. Customers may also return unopened/unexpired affected HLS Sets by contacting Customer Support. If an affected Emergency Priming Line is already in use, the firm advises it should remain in use due to increased potential risk of stopping this emergency procedure. Users should monitor the patient closely for signs of infection. Contact Getinge Customer Support: (888) 943-8872 (press option 2), Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. (Eastern Time Zone). ***Updated 3/21/24*** Additional lots identified. Firm notified affected customers on March 21, 2024.