SCC Soft Computer, Softlab Laboratory information System with DORA versions: 2.0.0.4 with Hot Fix 1.2808, 1.2809, 1.2956, or 1.3476; 2.0.0.5; 2.0.0.6; 2.0.0.7 with Hot Fix 1.6595; and 2.1.0.3.

This recall has been terminated (originally issued November 23, 2011).

Company
Scc Soft Computer
Recall Initiated
March 22, 2011
Posted
November 23, 2011
Terminated
October 29, 2012
Recall Number
Z-0278-2012
Quantity
12
Firm Location
Clearwater, FL
Official Source
View on FDA website ↗

Reason for Recall

Potential for patient data to be sent to the wrong patient record. A software change affected how the Softlab handles recovery from certain types of table access conflicts. When such conflicts occur, this function has the potential to cause faulty writing to the database.

Distribution

Sold throughout US and Canada.

Lot / Code Info

Softlab with DORA versions: 2.0.0.4 with Hot Fix 1.2808, 1.2809, 1.2956, or 1.3476; 2.0.0.5; 2.0.0.6; 2.0.0.7 with Hot Fix 1.6595; and 2.1.0.3.

Root Cause

Software change control

Action Taken

Clients were notified through SCC Soft Computer's proprietary Task Management System On 03/22/201. All clients have received a hot fix that corrected the issue or made the hosparam change that prevents the issue.