Terumo Advanced Perfusion Assembly Pump Guts, 6 Inch, Catalog Number 801798 - is indicated for use for up to 6 hours in the extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures, when used by a qualified medical proffessional who is experienced in the operation of this or similar equipment. The roller pump is used to move fluids through the cardiopulmonary bypass circuit and can be used for a number of applicatinos including arterial blood pumping, cardioplegia delivery, suction, and venting.

Company
Terumo Cardiovascular Systems Corporation
Recall Initiated
January 10, 2020
Terminated
May 12, 2020
Recall Number
Z-1110-2020
Quantity
16
Firm Location
Ann Arbor, MI

Reason for Recall

These roller pumps were assembled with pump shafts that may fail at the location of the thread where a locknut is installed. The failure is associated with a noise that is atypical of roller pump operation. The noise is caused when an unsecured shaft end contacts the rotating pulley of the roller pump. Prolonged use of a roller pump in this condition could lead to an eventual pump stop, resulting in the inability to create pump flow due to pulley loosening and drive belt damage.

Distribution

Worldwide distribution - US Nationwide distribution in the states of CT, FL, LA, MA, MI, NY, and TX. Countries of Hong Kong, India, South Korea, and Taiwan.

Lot / Code Info

Serial Numbers: 882298 882299 882300 882752 882753 882794 882795 882796 882797 882798 882799 885547 885548 885549 885550 885553

Root Cause

Nonconforming Material/Component

Action Taken

Terumo began notifying customers via telephone to schedule the repairs on 1/2/20. Urgent Medical Device Correction notification letters dated 1/6/20 were sent to customers.