MicroVention Terumo Microplex 10, Platinum Coil System, Endovascular Embolization Coil, HyperSoft Helical, Sterile, Rx, REF numbers 100101HS-V, 100102HS-V, 100103HS-V, 100104HS-V, 100151HS-V, 100152HS-V, 100153HS-V, 100154HS-V, 100201HS-V, 100202HS-V, 100203HS-V, 100204HS-V, 100206HS-V, 100254HS-V, 100256HS-V, 100304HS-V, 100306HS-V, 100308HS-V, 100406HS-V, 100408HS-V, and 100508HS-V. Product Usage: The MCS is also intended for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolizations in the peripheral vasculature. The firm name on the label is MicroVention, Inc., Tustin, CA.

Company
Microvention Inc.
Recall Initiated
November 22, 2019
Terminated
January 26, 2022
Recall Number
Z-1407-2020
Quantity
5,777 units
Firm Location
Aliso Viejo, CA

Reason for Recall

The devices may be missing the implant coil.

Distribution

All distribution was OUS. The countries receiving affected product are: Albania, Austria, Belarus, Belgium, Brazil, Bulgaria, China, Cyprus, Czech Republic, Denmark, Egypt, Finalnd, France, Georgia, Germany, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea, Latvia, Lebanon, Malta, Morocco, Netherlands, Norway, Poland, Portugal, Republic of Serbia, Romania, Russian Federation, Slovak Republic, Slovenia, South Africa, Spain, Sweden, Switzerland, Syria, Turkey, Ukraine, United Arab Emirates, and United Kingdom.

Lot / Code Info

Catalog/lot numbers:  100101HS-V - 1903045UH, 1903065UH, 1903115UH, 1903135UH, 1903155UH, 1903185UH, 1903205UH, 1903225UH, 1903225ZH, 1904085UH, 1904105UH, 1904175UH, 1904245UH, 1905025UH, 1905085UH, 1905105ZH, 1905135UH, 1905175UH, 1905205UH, 1905205ZH, 1905225UH, 1905225ZH, 1903135WH, 1904175WH, 1905085WH, 1905105PH, and 1905135WH;  100102HS-V - 1903115UH, 1903135UH, 1903155UH, 1903185UH, 1903205UH, 1903225UH, 1903225ZH, 1905085UH, 1905105UH, 1905105ZH, 1905135UH, 1905155UH, 1905155ZH, 1905175UH, 1905205UH, 1905205ZH, 1905225UH, 1905225ZH, 1904035WH, and 1905155WH;  100103HS-V - 1903115UH, 1903135UH, 1903155UH, 1903185UH, 1903205UH, 1903225UH, 1904105UH, 1904155UH, 1904175UH, 1904225UH, 1904245UH, 1905025UH, 1905025ZH, 1905035UH, 1905035ZH, 1905065UH, 1905085UH, 1905105UH, 1905105ZH, 1905135UH, 1905155UH, 1905155ZH, 1905175UH, 1905205UH, and 1905085WH;  100104HS-V - 1903115UH, 1903135UH, 1903225UH, 1904085UH, 1904105UH, 1904155UH, 1905085UH, 1905105UH, 1905105ZH, 1905135UH, 1905155UH, 1905155ZH, 1905175UH, 1905205UH, 1904175WH, 1904225WH, and 1905155WH;  100151HS-V - 1905025UH, 1903135TH, 1903085WH, 1903205WH, 1904015WH, 1904265WH, 1904295WH, 1905085WH, 1905155WH, and 1905225PH;  100152HS-V - 1903155UH, 1903185UH, 1903205UH, 1904085UH, 1904105UH, 1904155UH, 1904225UH, 1904245UH, 1905025UH, 1905025ZH, 1905035UH, 1905035ZH, 1905065UH. 1905085UH, 1905105UH, 1905135UH, 1905155UH, 1905155ZH, 1905175UH, 1905205ZH, 1903065NH, 1904175TH, 1904225NH, 1905065NH, 1903015WH, 1903065WH, 1903185WH, 1904015WH, 1904035WH, 1904055WH, 1904085WH, 1904155WH, 1904225WH, 1904245WH, 1904265WH, 1904295WH, 1905025WH, 1905035PH, 1905035WH, 1905065WH, 1905085WH, 1905155WH, and 1905225WH;  100153HS-V - 1903225UH, 1905105UH, 1905135UH, 1905135ZH, 1905155UH, 1905175UH, 1905205UH, 1905225UH, 1905225ZH, 1903015NH, 1903065WH, 1903085WH, 1903205WH, 1904015WH, 1904035WH, 1904055WH, 1904155WH, 1904175WH, 1904225PH, 1904225WH, 1904245WH, 1904265WH, 1905025WH, 1905035PH, 1905065WH, 1905085WH, 1905155WH, 1905205PH, 1905225PH, and 1905225WH.  100154HS-V - 1903155UH, 1903185UH, 1903205UH, 1903225UH, 1905085UH, 1905105UH, 1905105ZH, 1905135UH, 1905155UH, 1905155ZH, 1905175UH, 1905205UH, 1905225UH, 1903115TH, 1903015WH, 1903045WH, 1903065WH, 1904035WH, 1904055WH, 1904225WH, 1904245WH, 1904265WH, 1904295WH, 1905035PH, 1905035WH, 1905065WH, 1905085WH, 1905155PH, 1905155WH, 1905175WH, and 1905205WH;  100201HS-V - 1903135UH  100202HS-V - 1904085UH, 1904105UH, 1904175UH, 1904245UH, 1904225NH, 1905065NH, 1905205NH, 1903085WH, 1904015WH, 1904055WH, 1904175WH, 1904245WH, 1904265WH, 1904295WH, 1905085WH, 1905225PH, and 1905225WH;  100203HS-V - 1903155UH, 1903185UH, 1903225UH, 1904225NH, 1905135NH, 1903015WH, 1903065WH, 1904035WH, 1904225WH, 1904245WH, 1904265WH, 1904295WH, 1905035WH, 1905155WH, and 1905225WH;  100204HS-V - 1904085UH, 1904105UH, 1904225UH, 1904245UH, 1905065UH, 1905085UH, 1905105UH, 1905135UH, 1905135ZH, 1905155UH, 1905175UH, 1905205UH, 1903045WH, 1903065WH, 1904015WH, 1904035WH, 1904155WH, 1904175WH, 1904225PH, 1904245WH, 1904265WH, 1904295WH, 1905025WH, 1905035WH, 1905085WH, 1905155WH, and 1905225WH;  100206HS-V - 1904085UH, 1904105UH, 1904155UH, 1904175UH, 1904265UH, 1905065UH, 1905085UH, 1905105UH, 1905105ZH, 1905135UH, 1905155UH, 1905175UH, 1905205UH 1905205ZH, 1905225UH, 1903065WH, 1903085WH, 1903185WH, 1904055WH, 1904175WH, 1904225WH, 1904245WH, 1904265WH, 1904295WH, 1905025WH, 1905035PH, 1905035WH, 1905065WH, 1905155WH, 1905175WH, and 1905225WH;   100254HS-V - 1903205WH, 1904015WH, 1904225WH, 1904265WH, 1904295WH, 1905025WH, 1905085WH, and 1905225PH;  100256HS-V - 1903065WH, 1903185WH, 1904015WH, and 1905155WH;  100304HS-V - 1903085NH, 1903135TH, 1904035TH, 1903065WH, 1905035PH, and 1905065WH;  100306HS-V - 1903115UH, 1903135UH, 1903155UH, 1903185UH, 1904085UH, 1903065WH, 1904035WH, 1904055WH, 1904155WH, 1904175WH, 1904225WH, 1904245WH, 1905035PH, 1905065WH, 1905205PH, and 1905225PH;  100308HS-V - 1903185UH, 1903225UH, 1905135UH, 1905175UH, 1905205UH, 1903085NH, 1903015WH, 1903065WH, 1903185WH, 1904015WH, 1904035WH, 1904055WH, 1904175WH, 1904225WH, 1904245WH, 1904265WH, 1904295WH, 1905025WH, 1905035PH, 1905035WH, and 1905065WH;  100406HS-V - 1903085WH, 1903185WH, 1904035WH, 1904225WH, and 1905205WH;  100408HS-V - 1905085UH, 1905105UH, 1905135UH, 1905155UH, 1903115TH, 1903015WH, 1903065WH, 1904265WH, 1904295WH, 1905085WH, 1905135WH, 1905155WH, and 1905205PH;  100508HS-V - 1903225TH, 1905205NH, 1903015WH, 1903085WH, 1903135WH, 1903185WH, 1904265WH, 1905105WH, and 1905205WH.

Root Cause

Process design

Action Taken

The recalling firm issued letters beginning 11/22/2019 dated 11/26/2019 via email using different flags according to the region the foreign consignee was located, i.e. 'URGENT: FIELD SAFETY CORRECTION ACTION" (for EMEA), "URGENT: MEDICAL DEVICE RECALL" (for Japan), or "FIELD SAFETY ADVISORY NOTICE" (for Rest of World).