Offset Reamer Handle, T5766 Product Usage: The reamer handles are re-usable surgical instruments used in arthroplasty surgery of the hip joints, where either total joint replacement or resurfacing surgery is performed using a reamer cutting instrument. The reamer handles are intended for use with a surgical driver (active device). The reamer handle is connected at the distal end to a reamer. The T5766 reamer handle is an offset design with a drive shaft mechanism that spins on roller bearings and is encapsulated in an inner clam shell housing that is constrained within a Radel tube and outer housing.
This recall has been terminated (originally issued October 19, 2016).
- Company
- Greatbatch Medical
- Recall Initiated
- October 19, 2016
- Terminated
- March 8, 2019
- Recall Number
- Z-1190-2017
- Quantity
- 1,441
- Firm Location
- Plymouth, MN
- Official Source
- View on FDA website ↗
Reason for Recall
On June 14, 2016, Greatbatch discovered a significant increase in the number of field complaints for the T5766 Offset Reamer Handle in which the U-joints of the drive chain component seizes during orthopedic procedures rendering the device non-functional. The 2016 failure rate for this failure exceeds the occurrence rate as defined in the DFMEA.
Distribution
Worldwide Distribution - US Nationwide in the states of IN and the countries of Canada, Austria, Germany, Japan.
Lot / Code Info
2481193, 2468380, 2599785, 2463309, 2468379, 2468398, 2468397, 2599516, 2671555, 2599517, 2670543, 2671558, 2734809, 2814828, 2814829, 2824496, 2858434, 2865404, 3030658, 2944397, 2977852, 2967214, 3000586, 3000590, 3030658, 3044689, 3071903, 3135658, 3088675, 3107504, 3188688, 3385844, 3188690, 3387046, 3397768.
Root Cause
Device Design
Action Taken
Consignee was sent via e-mail on 10/16/2016 a Greatbatch "Urgent Medical Device Recall" letter dated 10/19/2016 The letter described the purpose of the letter, Reason for Notification, Risk to Health, Recommendation, Instruction to Zimmer. Advised consignee to examine their inventory, quarantine the product and provide a list of the affected countries where the product was distributed. Also requested consignee to complete the Customer Notification Response form and return to Greatbatch at FieldActionCenter@Greatbatch.com. For questions contact your local Greatbatch Medical representative, Monday through Friday via telephone at 1-763-951-8235 8:00 AM (CT) to 5:00 PM (CT), or e-mail us at FieldActionCenter@Greatbatch.com .