Terumo 100/120V AC, Advanced Perfusion System Platform (APS) for use in cardiopulmonary bypass procedures. Catalog Number 801763
- Recall Initiated
- August 28, 2018
- Terminated
- April 23, 2020
- Recall Number
- Z-3242-2018
- Quantity
- 47 units
- Firm Location
- Ann Arbor, MI
Reason for Recall
Screws on a Terumo Advanced Perfusion System 1 Power Manager Board (the base unit circuit board) were not torqued to specification. It is possible that a screw which has not been properly torqued can go undetected during manufacturing testing and may loosen over time, resulting in the potential for an electrical short in the system.
Distribution
Nationwide Foreign: Mexico, BELGIUM, CANADA, India, Indonesia, Singapore, South Korea , Vietnam
Lot / Code Info
Serial Numbers: 0386 1481 1497 1509 1526 1529 1531 1532 1535 1536 1537 1538 1539 1540 1541 1542 1543 1544 1545 1546 1548 1549 1550 1551 1552 1553 1554 1555 1556 1557 1558 1561 1562 1563 1564 1565 1566 1570 1597 1612 1630 1631 1640 1641 1649 1547 1560 1559 1567 Spare Parts: 870509 ASSEMBLY POWER MANAGER POD APS S/N: 03130 , 03308
Root Cause
Process control
Action Taken
Terumo CVS issued letter on 8/2/818 via Fed'X 2 day air to inform users of the problem for the Power Manager boards with unverifiable or out of specification screws, potential hazard and field correction. A Terumo Field Service Representative will contact users to schedule the field correction activities. Terumo CVS Customer Service: 1.800.521.2818, Monday Friday, 8 a.m. 6 p.m. ET