CELL-DYN Emerald Cleaner, product number 09H46-02. 960mL. IVD; Hematology: The CELL-DYN Emerald Cleaner is designed for use on the CELL-DYN Emerald System, which is an automated hematology analyzer designed for in-vitro diagnostic use in clinical laboratories. The CELL-DYN Emerald Cleaner is an enzymatic cleaner used to clean the measurement system and the fluidics.

Company
Abbott Laboratories
Recall Initiated
March 24, 2016
Posted
April 19, 2016
Terminated
January 4, 2017
Recall Number
Z-1522-2016
Quantity
10,364 plus 9410 additional units for recall expansion.
Firm Location
Santa Clara, CA

Reason for Recall

Three lots of CELL-DYN Emerald Cleaner may have occurrences where the analyzer generates Quality Control Out of Range Low for parameters RBC and PLT.

Distribution

Worldwide Distribution-US (nationwide) including Puerto Rico and the countries of Austria, Bosnia & Herze, Cameroon, Egypt, Germany, Indonesia, Iraq, Lebanon, Mali, Montenegro, Netherlands, Philippines, Tanzania, Thailand, UAE, New Zealand, Malaysia, Mexico, South Korea, Jamaica, Turks & Caicos, Italy, Canada, Vietnam, Singapore, Argentina, Colombia, France, Paraguay, Australia, Uruguay, Bahrain, Belgium, Bulgaria, Croatia, Cyprus, Czech Republic, Ethiopia, Finland, Greece, Switzerland, United Kingdom, Brazil, Hong, Kong, Hungary, India, Ireland, Israel, Kazakhstan, Kenya, Kuwait, Latvia, Lithuania, Luxembourg, Turkey, Jamaica, St. Martin, Nigeria, Oman, Pakistan, Philippines, Poland, Qatar, Romania, Russia, Saudi Arabia, Serbia, Slovakia, Slovenia, Spain, Guatemala, Peru, and Uruguay.

Lot / Code Info

Lot 6853, Expiration date: 31 JUL2017; Lot 6901, Expiration date: 31JUL2017; Lot 6953, Expiration date: 30SEP2017;   May 4, 2016 - Firm advised SAN-DO that it has expanded the recall to include three additional lots:  Lot 6991, Expiration date: 30NOV2017; Lot 7024, Expiration date: 31DEC2017; Lot: 7027, Expiration date: 31DEC2017.  June 28, 2016 - firm added an additional 4 lots to recall: Lot 7044, expiration date 31JAN2018; Lot 7082, expiration date 28FEB2018; Lot 7110, expiration date 31MAR2018; Lot 7119, expiration date 30APR2018.

Root Cause

Under Investigation by firm

Action Taken

All customers who were shipped the affected lots were sent a Product Recall letter on March 24, 2016 via Federal Express. Letters were titled Product Recall - Immediate Action Required. An additional four lots were discovered to be affected by the recall and the firm send letters to customers affected by these lots on June 27, 2016