Emerald Diluent Tubing, List number 09H50-01, accompanying CELL-DYN Emerald instrument, List number 09H39-01. The firm name on the label is Abbott Laboratories, Abbott Park, IL.

Company
Abbott Laboratories
Recall Initiated
May 9, 2019
Terminated
October 7, 2020
Recall Number
Z-0173-2020
Quantity
22 analyzers with tubing
Firm Location
Santa Clara, CA

Reason for Recall

The diluent tubing used on the instrument appears cloudy or to have a powder/film on the inner and outer surfaces which may result in falsely elevated platelet results.

Distribution

Distribution was made to GA, LA, MA, NC, NY, SC, and TX. There was no military/government distribution. Foreign distribution was made to Australia, Germany, Greece, Honduras, Indonesia, Italy, and Serbia.

Lot / Code Info

Tubing lot #2018-10-16; Instrument serial numbers 007689, 007690, 007691, 007692, 007705, 007706, 007707, 007708, 007709, 007728, 007694, 007695, 007697, 007766, 007767, 007768, 007765, 007710, 007699, 007700, 007701, and 007703.

Root Cause

Nonconforming Material/Component

Action Taken

The recalling firm issued letters on 5/9/2019 to the consignees who had the installed instruments with the affected serial numbers. The letters were issued via FedEx priority overnight express.