CELL-DYN Ruby System: List Number: 08H6701; Hematology: The CELL-DYN Ruby System is a multi-parameter automated hematology analyzer designed for in-vitro diagnostic use in clinical laboratories

Company
Abbott Laboratories
Recall Initiated
February 13, 2017
Posted
March 9, 2017
Terminated
March 6, 2018
Recall Number
Z-1438-2017
Quantity
116
Firm Location
Santa Clara, CA

Reason for Recall

The Pump Relay Printed Circuit Board Assembly (PCBA) used in the CELL-DYN Ruby instrument may prematurely fail and lead to instrument stoppage

Distribution

US and Afghanistan, Brazil, Egypt, Tanzania, Hungary, Iraq, Kenya, Netherlands, Russia, Romania Vietnam, , United Kingdom, Saudi Arabia, Philippines, Lebanon, Mexico, Japan, Jordan, India, Indonesia, Turkey, UAE, Gaza & Jericho, Germany, Bulgaria, Chile, Argentina, Luxembourg.

Lot / Code Info

Affected serial numbers:  70002BG 70026BG 70050BG 70076BG 70105BG 70141BG 70003BG 70028BG 70051BG 70077BG 70106BG 70142BG 70004BG 70029BG 70052BG 70079BG 70107BG 70144BG 70005BG 70030BG 70054BG 70082BG 70113BG 70146BG 70006BG 70031BG 70056BG 70084BG 70114BG 70147BG 70007BG 70032BG 70058BG 70086BG 70119BG 70149BG 70008BG 70033BG 70059BG 70087BG 70120BG 70164BG 70009BG 70034BG 70061BG 70088BG 70124BG 70165BG 70010BG 70035BG 70063BG 70089BG 70125BG 70166BG 70011BG 70036BG 70064BG 70090BG 70127BG 70167BG 70012BG 70037BG 70065BG 70091BG 70129BG 70170BG 70013BG 70038BG 70067BG 70092BG 70130BG 70171BG 70014BG 70039BG 70068BG 70093BG 70131BG 70178BG 70016BG 70040BG 70069BG 70094BG 70132BG 70181BG 70018BG 70042BG 70070BG 70095BG 70133BG 70183BG 70019BG 70043BG 70071BG 70096BG 70134BG 70189BG 70020BG 70044BG 70072BG 70099BG 70136BG 70021BG 70045BG 70073BG 70100BG 70137BG 70024BG 70046BG 70074BG 70103BG 70138BG 70025BG 70047BG 70075BG 70104BG 70140BG.  UPDATE: 2-28-17 - five additional serial numbers added: 55899BG,  55900BG,  55901BG,  55902BG  55904BG.

Root Cause

Nonconforming Material/Component

Action Taken

Product correction letters were sent to all affected customers on February 13, 2017 by traceable method. Letters provided information on the reason for the correction with all affected serial numbers listed. Abbott will be contacting each location to schedule replacement of the PCBA Pump Relay Board. Update: five additional serial numbers were added. Correction letters were sent to new consignees - 3 in Brazil and 2 in Mexico.